Posted : 05/01/1999
How Much Is Enough?
Measuring Output Is Crucial For Tanning Success
by Judie Gaillard
Whether you are a seasoned tanning salon owner or the new kid on the
block, understanding lamps and how they work is crucial to continued success. In addition
to understanding which lamps are compatible with your tanning equipment, salon operators
must have a firm understanding of tanning lamp care and output so they can determine when
it is time to relamp equipment, as well as knowledge of maintenance factors that can
affect a lamp's performance.
The Numbers Game
Every salon operator knows that the tanning lamp is probably the single most important
variable in the tanning equation. In fact, more and more salon owners tell us that their
customers are becoming more educated about lamps and asking questions about strengths and
output.
Although tanning lamps may seem simple to order and replace, they are complex
components that figure heavily into the overall tanning process. Each lamp manufacturer
has its own definition of useful life for its product, and each tanning salon has its own
set of operating conditions. Although both these points are not very helpful in answering
the question of when you should replace your lamps, there are some guidelines that can
help determine the right point in time.
When examining the output performance of a lamp, as a rule, about 20 percent to 25
percent of the electrical input will be emitted as ultraviolet radiation. However, this
only is true for new lamps that are in optimal operating conditions--meaning that the
electrical conditions such as supplied voltage, lamp current and wattage meet the
specifications of the lamp manufacturer.
Ballasts and starters (if used) also play an important role. The two major factors
responsible for the actual performance of a tanning unit are the aging of the lamps over
time and equipment characteristics of the tanning unit.
Any radiation source loses a certain degree of power the longer it is in operation. For
example, the reduction in power hardly is noticed in the general lighting of private
households. In general, incandescent or fluorescent lamps are not replaced until they stop
functioning.
Unfortunately, this is not as simple with the UV lamps used for indoor tanning. Tanning
units are expected to meet certain requirements with respect to their tanning
effectiveness over a given period of time. With increased use, they are no longer able to
meet such expectations because of a marked performance decline, even though they are still
in full working order from an electrical point of view. With tanning lamps, the
recommended "useful life" is considerably shorter than the electrical life.
Normally, manufacturers give recommendations on the "useful life" of their
lamps, but these recommendations only can be used as a guide because there is no clear and
official definition of the term "useful life." Each manufacturer can make its
own definition. Additionally, different operating conditions, as well as equipment-related
factors, have in certain cases a considerable effect on the actual useful life of a
tanning lamp.
To Replace Or Not To Replace
General consensus indicates that the best and most reliable means of determining when
lamps should be replaced is through the correct use of a UV meter. The basic rule of thumb
is when the output level has dropped to 70 percent to 75 percent of what it was when the
lamps were new, the lamps should be replaced.
Generally, lamps show the highest decline in their output performance during the first
100 hours of operation. For this reason, some lamp manufacturers define their lamp's
nominal output performance after 100 hours of operation.
The drop in power between zero and 100 hours can amount to about 10 percent to 20
percent, depending on the properties of the phosphor used inside the lamp. Most of the
published maintenance curves were obtained with "free burning" lamps under
optimal operating conditions. Therefore, the actual decline in the output of lamps in
tanning beds often differs significantly when taken from the salon environment.
There are two different classes of phosphor available for tanning lamps. One type is
used in standard tanning lamps and shows a stronger decline in performance, recommending
replacement after approximately 300 hours to 500 hours of operation. Another type of
phosphor is considerably more expensive and largely used in the professional market of
tanning lamps. The gradual decline in performance of these lamps extends their useful life
up to 800 to 1,000 hours.
Measuring UVA irradiance is often part of a daily routine for tanning salon operators.
For reasons of practicality and cost, the UVA measuring instrument preferred for this
purpose is generally an inexpensive pocket-sized type. For financial reasons and customer
satisfaction, there should be the best possible assurance that the salon owner can depend
on the readout from his or her UV meter.
A UV meter can be a great tool when used properly; however, all too often salon owners
are under the misconception that it is an exact measurement for output. If salon owners
choose to rely on the UV meter, they should be sure to follow the same format for
measuring output which is outlined below.
- Take an initial reading upon the installation of new lamps.
- Make sure the measuring conditions always are kept the same. Measure at the same
location of your tanning unit, at the same distance and observe the same electrical
conditions for each single measurement.
- Make sure that the acrylics and reflectors are clean. It is imperative that acrylics are
changed according to the manufacturer's specifications.
- Wait for your measurement until your tanning unit is in a state of "thermal
equilibrium," which takes about 20 minutes to 30 minutes after turning it on.
If these steps are followed, measurement readings for comparative purposes will be
obtained. Pocket-sized UVA meters are used to measure UVA values. Although these meters
only register UVA levels, it can be assumed that the UVB values will diminish by
approximately the same relative amount.
Factors That Affect Output
The total operating time of the lamps is probably the most important factor determining
output performance. However, some additional factors stem from the brand of tanning unit
used and how well it is functioning.
Equipment-related factors such as dirty or dusty reflectors will reduce power. These
issues can be avoided by regularly cleaning your equipment. Other factors include the
permeability of the acrylic cover and the cooling system.
Acrylics
Acrylic care is probably the most overlooked element of tanning bed
maintenance. It must be noted that it is essential to optimize lamp performance through
the use of acrylics that allow for proper transmission of UV energy.
Make sure to check that the reduction of irradiance is not due to insufficient
permeability or penetration of the acrylic sheet. Simply use your UVA meter and compare
the readings made with and without the acrylic sheet. If UVA readings obtained with the
acrylic sheet on the tanning unit are more than 20 percent lower compared to those without
the sheet, you probably have an aging acrylic sheet which acts as a filter.
Additionally, yellowing of the cut edges on acrylics indicates it is time to replace
the acrylic sheet. It is important that the acrylics be changed with the manufacturer's
specifications.
Cooling Systems
As is generally known, a lamp can provide maximum power only if it is
operated within an optimum temperature range that is approximately 108 degrees Fahrenheit.
If the temperature is too low or too high, it will lead to a drop in output. To check
whether the cooling system of your unit conforms to the optimal operating conditions of
the lamps, the following procedure can be recommended.
During the warm-up phase of up to approximately 30 minutes after switching on your
unit, constantly record the UVA irradiance by use of a UVA meter. The best readings will
be obtained in the center of the tanning area. If the reading becomes steady at maximum
UVA values, it can be assumed that the cooling of the unit is proportioned correctly.
If, on the other hand, the measured values pass through a maximum before settling at a
lower reading, the flow of cooling air is probably too weak. If there is a constant rise
in the readings throughout the entire warm-up phase, but without a noticeable maximum
being obtained, this should be interpreted as a sign that the unit may be over-cooled.
Such problems can be adjusted if the unit allows you to change the cooling airflow. If
not, contact the manufacturer of the unit.
Lampholders
Salon operators also should make sure their lampholders are in proper
working condition since they support fluorescent lamps and provide electrical connection.
There are numerous types of fluorescent lampholders available for the different types of
lamp bases.
The most frequently used connector for preheat and rapid start bi-pin lamps is the
twist-turn type. Spring pressure push-pull lampholders also are used for bi-pin base
lamps. For single lamp ballasts and dimming ballasts, special circuit-interrupting bi-pin
lampholders are available. In the case of some single pin lamps, support comes in the form
of a high-voltage lampholder with a low-voltage circuit-interrupting feature. This type of
connector prevents voltage from being applied to the pins of the lamp until it is secured
firmly in both lampholders, reducing the possibility of shock when installing the lamps.
Ballasts & Starters
Ballasts and starters also contribute to lamp performance. Ballasts
regulate the flow of electricity into the lamp and they are rated according to the wattage
of the lamp for which they are designed. In other words, a 100-watt lamp requires a
100-watt ballast.
Traditionally, ballasts are either electronic or magnetic. The electronic ballasts have
the advantage of being lighter than their magnetic counterparts. Magnetic ballasts, while
heavier, are more resistant to heat and have fewer parts to fail.
Because of their weight, the placement of ballasts in the top portion of a bed is
sometimes used for balance, making the top easier to open and close. If a ballast needs to
be replaced, the new one should be attached in the same place as the old, to avoid
altering the balance of the canopy.
Separate starters most often are found in beds of European origin. Starters are the
spark plugs of tanning beds. When current is applied to the lamp circuit, the starter
sends a high-voltage pulse through the electrodes, warming up the lamp and initiating the
arc between the electrodes. If starter output is inconsistent or incorrect, you may notice
blackening of the lamp ends.
|